Muffle furnaces with hinged or lifting door

+ Free Shipping

The Nabertherm L 3/11 – LT 40/12 muffle furnace range is perfect for daily laboratory use. Excellent performance, modern design and a high degree of reliability are the hallmarks of this model range. Muffle furnaces are supplied with a hinged or lifting door.

The Nabertherm L 3/11 – LT 40/12 muffle furnace range is perfect for daily laboratory use. Excellent performance, modern design and a high degree of reliability are the hallmarks of this model range. Muffle furnaces are supplied with a hinged or lifting door.

  • Maximum temperature 1100 ° C or 1200 ° C;
  • Heating the oven from both sides with ceramic burners (for models L 24/11 – LT 40/12, heating from three sides);
  • Ceramic heating plates with an integrated heating element, protected against splashes and fumes (easy to replace);
  • Vacuum hardened high strength fiber module;
  • Muffle furnace body made of structured stainless steel sheets;
  • Double wall muffle furnace body for low external temperatures and high stability;
  • Muffle furnaces are available as an option with either a hinged door (L), which can be used as an additional loading and unloading area, or a lifting door (LT), with the hot side facing away from the operator;
  • Adjustable intake air opening in the muffle furnace door;
  • Exhaust hole in the rear wall of the furnace;
  • Quiet operation of the muffle furnace heating system with semiconductor relays.

Additional equipment for muffle furnaces:

  • Chimney, chimney with fan or catalytic converter;
  • Adjustable temperature limiter with adjustable shutdown temperature for thermal protection class 2 according to EN 60519-2 as overheating protection for ovens and products;
  • Shielding gas supply at the rear of the furnace;
  • Manual or automatic gas supply system.
Technical data for Nabertherm hinged door muffle furnaces
Furnace model T max , ° C Internal dimensions, mm Volume, l External dimensions, mm power, kWt Power supply Weight, kg Minutes to T max
Sh D IN Sh D IN
L 3/11 1100 160 140 100 3 380 370 420 1,2 1-phase twenty 60
L 5/11 1100 200 170 130 5 440 470 520 2.4 1-phase 35 60
L 9/11 1100 230 240 170 9 480 550 570 3.0 1-phase 45 75
L 15/11 1100 230 340 170 15 480 650 570 3.6 1-phase 55 90
L 24/11 1100 280 340 250 24 560 660 650 4.5 3-phase 75 95
L 40/11 1100 320 490 250 40 600 790 650 6.0 3-phase 95 95
L 3/12 1200 160 140 100 3 380 370 420 1,2 1-phase twenty 75
L 5/12 1200 200 170 130 5 440 470 520 2.4 1-phase 35 75
L 9/12 1200 230 240 170 9 480 550 570 3.0 1-phase 45 90
L 15/12 1200 230 340 170 15 480 650 570 3.6 1-phase 55 105
L 24/12 1200 280 340 250 24 560 660 650 4.5 3-phase 75 110
L 40/12 1200 320 490 250 40 600 790 650 6.0 3-phase 95 100
Technical Data for Nabertherm Lift Door Muffle Furnaces
Furnace model T max , ° C Internal dimensions, mm Volume, l External dimensions, mm power, kWt Power supply Weight, kg Minutes to T max
Sh D IN Sh D IN*
LT 3/11 1100 160 140 100 3 380 370 420 + 165 1,2 1-phase twenty 60
LT 5/11 1100 200 170 130 5 440 470 520 + 220 2.4 1-phase 35 60
LT 9/11 1100 230 240 170 9 480 550 570 + 290 3.0 1-phase 45 75
LT 15/11 1100 230 340 170 15 480 650 570 + 290 3.6 1-phase 55 90
LT 24/11 1100 280 340 250 24 560 660 650 + 335 4.5 3-phase 75 95
LT 40/11 1100 320 490 250 40 600 790 650 + 335 6.0 3-phase 95 95
LT 3/12 1200 160 140 100 3 380 370 420 + 165 1,2 1-phase twenty 75
LT 5/12 1200 200 170 130 5 440 470 520 + 220 2.4 1-phase 35 75
LT 9/12 1200 230 240 170 9 480 550 570 + 290 3.0 1-phase 45 90
LT 15/12 1200 230 340 170 15 480 650 570 + 290 3.6 1-phase 55 105
LT 24/12 1200 280 340 250 24 560 660 650 + 335 4.5 3-phase 75 110
LT 40/12 1200 320 490 250 40 600 790 650 + 335 6.0 3-phase 95 100
* incl. open lift door

Controller B410

  • Easy to install the program
  • Black and white LC screen, high contrast
  • Display text messages (2 lines)
  • Input data with 4 keys (on/off/pause, menu, back, info)
  • Stores 5 programs, each program can be adjusted in 4 steps
  • Step of temperature adjustment 1 o C
  • Accuracy ± 1°C
  • Choice of °C/°F
  • Display on the menu parameters:
  • Working timer
  • Count kWh
  • Show error history
  • Output of control values
  • Auto-optimization feature
  • Calibration of measuring range with selection of up to 10 points
  • NTLog USB port for data retrieval and processing
  • In case of power failure, the following modes can be selected:
  • Automatically resume running according to the program right before the power failure
  • Automatically continue running according to the program right before the power failure if the power failure time is < 2 minutes, if more than that, do not continue running.
  • Automatically continue running according to the program right before the power failure if the temperature does not drop lower than 50oC, if it is lower, do not continue running
  • Do not continue to run the old program when restoring the power

Click Our Social Network Links

Best seller & Supplier in Bangladesh

Al-Arafa Scientific & Surgical Co Ltd.

Click for inquiry & purchase, Related products

Google, Facebook, LinkedIn, Twitter, Instagram, YouTube

You may also like…

Shopping Cart
Seraphinite AcceleratorOptimized by Seraphinite Accelerator
Turns on site high speed to be attractive for people and search engines.